Nanolive 3D Cell Explorer User Manual (Analysis & Translation)
Product: Nanolive 3D Cell Explorer (incl. 3D Cell Explorer-fluo)
Original manufacturer: Nanolive SA (Lausanne, Switzerland) β www.nanolive.ch
Structure: 9 chapters / 60 pages / 16 selected figures
1. What is it? π¬
The 3D Cell Explorer is Nanolive's label-free live-cell 3D tomographic microscope based on Holographic Tomography. A Class I laser (Ξ»=520 nm, 0.2 mW/mmΒ² sample exposure, non-phototoxic) passes through live cells, and the refractive index (RI) distribution is recorded to reconstruct the cell's 3D internal structure.
1.1 Three major differences from conventional microscopes
| Dimension | Conventional microscope | 3D Cell Explorer |
|---|---|---|
| Labeling | Requires fluorescence/staining | Fully label-free (RI-based) |
| Dimensionality | 2D or confocal 3D (sectioning required) | Single-shot 3D reconstruction (0.5 fps) |
| Phototoxicity | Laser/mercury/LED damages cells | Class I laser, 0.2 mW/mmΒ², no phototoxicity for long-term live-cell imaging |
| Data | Grayscale + fluorescence channels | 3D RI volume + digital staining (physical-parameter-based, quantifiable) |
1.2 Two versions
- 3D Cell Explorer: Pure RI holographic tomography (bright-field mode for sample search only)
- 3D Cell Explorer-fluo: Adds 3-channel fluorescence (up to 4 filter configurations) on top of RI β same FoV shows both "structure" and "chemistry" with co-localization in one acquisition
1.3 Package contents (16 items)
| # | Item | Qty |
|---|---|---|
| 1 | 3D Cell Explorer main unit | 1 |
| 2-5 | USB 2.0 + USB 3.0 + power + IEC cables | 1 each |
| 6 | Cleaning kit | 1 |
| 7 | Reusable dust cover | 1 |
| 8 | WEEE certificate | 1 |
| 9 | Microscope objective (MO) protective cap | 1 |
| 10 | BNC cable (fluo only) | 1 |
| 11-16 | CoolLED illumination system (fluo only): LED source + DC power + IEC cable + USB 2.0 + Liquid Light Guide (LLG) + 3 excitation filter holders | 1 each |
2. Instrument overview π·
2.1 Three views
Fig 1 β Three external views showing rotating arm, XY stage, scanning head, camera, and bottom interface layout.
2.2 Key components terminology
Fig 2 β 7 core components: Scanning head / Rotating arm / XY stage / Z knob (focus) / Microscope objective (60X NA 0.8) / Camera (CMOS 1024Γ1024 @ 165 fps) / Peripheral mirror (for 3D tomography)
2.3 Fluo version and CoolLED
Fluo version: same external shape as base, with additional LLG liquid light guide port and BNC trigger on the back.
CoolLED pE-800 series: supports 3-4 excitation filters simultaneously (DAPI/FitC/TritC/Cy5), switch time <100 ΞΌs, lifetime >20,000 h per channel.
3. Technical specifications π
3.1 Base version
| Spec | Value | Note |
|---|---|---|
| Dimensions (WΓDΓH) | 380Γ170Γ445 mm | Desktop |
| Weight | 8 kg | Single-person portable |
| Operating temp | 20-30Β°C / 100-240 VAC / 50-60 Hz / 1.0-0.5 A | Temperature control critical |
| Camera | CMOS 165 fps, 1024Γ1024 px, USB 3.0 | Native sampling |
| Objective | 60X, dry (non-oil) | Maintenance-free |
| Laser | Class I (Ξ»=520 nm, 0.2 mW/mmΒ²) | No goggles needed |
| FoV | 85Γ85Γ30 Β΅m | Per 3D reconstruction |
| Lateral resolution | 200 nm | Sub-cellular |
| Axial resolution | 400 nm | |
| Tomography frame rate | 0.5 fps (one full 3D every 2 s) | |
| Self-adjustment time | <90 s | Depends on optical thickness |
3.2 Fluo version + CoolLED
| Spec | Holographic tomography | Fluorescence |
|---|---|---|
| Dimensions | 380Γ170Γ445 mm | 77Γ186Γ162 mm (CoolLED controller) |
| Weight | 8 kg | 4 kg |
| Operating temp | 20-30Β°C | 5-35Β°C |
| Camera | USB3.0 CMOS Sony IMX174 (QE 70% @545 nm, dark noise 6.6 eβ», DR 73.7 dB) | β |
| FoV | 90Γ90Γ30 Β΅m | 90Γ90 Β΅m |
| Lateral res | 200 nm | 400 nm |
| Channels | up to 7 simultaneous (RI digital stains) | 3 fluorescence (3 std / 4 max) |
| Frame rate | 0.5 fps | 3 fps/channel |
| Filters | β | Std: FitC+TritC+DAPI or FitC+TritC+Cy5; Full 4: TritC and Cy5 interchangeable on channel 3 |
3.3 Computer requirements (mandatory)
β οΈ GPU is the hard threshold β without NVIDIA discrete 8 GB, STEVE cannot run full version.
| Item | Minimum |
|---|---|
| CPU | 2.0 GHz quad-core or faster |
| RAM | 16 GB DDR3 |
| Storage | 2 GB system for STEVE + separate 1 TB data drive (continuous @0.5 fps = 1-2 GB/h) |
| GPU | NVIDIA discrete 8 GB VRAM (4 GB insufficient) |
| USB | At least 1 USB 3.0 port (for camera) |
| OS | Windows 10 64-bit + latest NVIDIA drivers + USB 3.0 drivers |
| Network | Internet for installation |
π‘ Supplier can provide a tested/validated PC/laptop list (manual P15) β request before purchasing.
4. First start-up β 9 physical installation steps π
4.1 Unboxing and positioning
Fig 3 β 16 package items. Fluo adds BNC cable and CoolLED system.
Step 1: Place on stable, flat, level, sturdy bench. Hold by base or sides β never lift by camera (P19 warning). Misuse damage not covered by warranty.
Step 2: Remove disposable plastic wrap and red vinyl MO cap.
4.2 CoolLED wiring (fluo only)
β οΈ Safety first β do not connect main power before LLG is installed.
Step 3: Place CoolLED near microscope, 200 mm clearance on each side for cooling (P22 Fig d).
Step 4: Insert 3 excitation filter holders (DAPI/FitC/TritC or Cy5) into CoolLED top slots, note orientation (P21).
Steps 5-6: Install LLG at CoolLED end.
Fig a β Remove red vinyl cap from LLG end.
Fig b β Insert LLG fully into CoolLED top port, secure with 1.5 mm hex key (don't over-torque β damages LLG head, reduces coupling).
Fig c/d β CoolLED with LLG installed, placed vertically with 200 mm clearance.
Step 7: Install LLG at microscope end.
Fig g β LLG other end into microscope side port, 3 set screws tightened in sequence. Any physical damage to LLG head causes irreversible loss of coupling efficiency.
π‘ LLG bend radius rule (P24 Fig i/j):
- Short-term bend R >40 mm
- Long-term bend R >75 mm
- 90 mm verified stable
- Never stretch, kink, or contact sharp objects β internal glass fibers break β reduced light output
Steps 8-9: Connect CoolLED power (DC PSU β CoolLED, plug flat side inward P25 Fig m) + USB 2.0 (CoolLED β PC) + BNC (CoolLED β microscope trigger).
4.3 Main unit power + USB
| Step | Action | Notes |
|---|---|---|
| 1 | USB 3.0: microscope side β PC | Tighten screws, must use PC's USB 3.0 port (not 2.0) |
| 2 | USB 2.0: microscope back β PC | Any USB port |
| 3 | Power β PSU β outlet | Confirm power switch off before plugging |
| 4 | Fluo: CoolLED power β IEC β outlet | CoolLED has no power switch β plugging in powers it |
4.4 Install STEVE software
Go to www.nanolive.ch/register (serial number on unit bottom / delivery note / quick start guide), download STEVE. Important: right-click STEVE.exe β "Run with graphics processor" β select "Nvidia high-performance" (integrated GPU = very slow).
5. Correct moving/transport β οΈ
P30-P37 β transport protection; customers often damage units by ignoring these.
5.1 Before moving
- Unplug all cables: power + USB 2.0 + USB 3.0 (fluo: add BNC + LLG)
- Use Z knob to raise stage to top position (otherwise hits MO during move)
- Lock stage with red set screw
- Fluo: cap LLG ends with red vinyl caps + cap MO with black vinyl cap
5.2 Stage locking
Left = high-damage transport position (stage not raised). Right = correct (Z knob up + red set screw locked).
Red set screw location. Must remove together when disassembling stage.
5.3 Carrying position
P37: hold bottom or sides β never lift by stage, rotating arm, or camera.
5.4 LLG end protection
- When not connected: cap both ends with red vinyl caps (P36 Fig o/p)
- LLG is precision optics, silica end-face scratches are irreversible
- Bend radius rule: short-term >40 mm / long-term >75 mm
- Clean: lens tissue + alcohol
6. Sample preparation π§«
P38 summary; detailed SOP in supplier's Nanolive Sample Preparation Protocol and Nanolive's Best Staining Practices (English).
6.1 Compatible containers
| Container | Use case |
|---|---|
| Glass coverslip | Fixed cells (after attachment) |
| FluoroDishβ’ | Live-cell long-term imaging (preferred) |
Not suitable: non-glass-bottom plastic plates, suspension cells in droplets (optical path through liquid column attenuates).
6.2 Key rules
- Extremely low light dose β suitable for long-term (>1 week) live-cell imaging
- Compatible with standard culture consumables β no special prep
- Set immersion medium RI correctly β STEVE prompts at acquisition start (water = 1.33, DMEM β 1.34, oily media must be checked)
7. Optical cleaning procedure π§Ή
P39-P40: cleaning is the only user-serviceable maintenance; other disassembly voids warranty.
General rules:
- Use Texwipeβ’ Microdenier or Alpha series swabs + ethanol/isopropanol only
- Wipe from center to edge in single direction (don't drag back and forth)
- Use a fresh swab each contact (avoid re-contamination)
- Clean most-exposed surfaces first (higher contamination probability)
7.1 Microscope objective (most frequent)
Fig 7 β MO is most contamination-prone (sample sits on it); needs regular cleaning.
7.2 Rotating arm scanning mirror (occasional)
Fig 8 β Scanning mirror on rotating arm: less exposed, infrequent cleaning.
7.3 Rotating arm center mirror (least frequent)
Fig 9 β Center mirror not exposed, occasional cleaning only.
π‘ When not in use, strongly recommend red vinyl cap on MO (P33) β prevention is easier than cleaning.
8. STEVE software β acquisition and digital staining π¨
P41-P55. STEVE = Software for Tomographic Explorer for Visualization and Evaluation.
8.1 Core features
| Module | Function |
|---|---|
| Microscope control | Single 3D, time-lapse 3D, bright-field assistant, fully auto self-adjust |
| 2D view | X-Y slice + user-defined staining overlay |
| 3D view | 3D volume rendering (voxel/surface) + screenshot + AVI recording |
| Digital staining | RI-based (not color) physical staining β quantifiable |
| Data export | raw/tiff/obj/png/jpeg + proprietary .vol/.vo1x |
| Annotation | Custom properties per measurement |
| Semi-auto update | New version auto-prompts; close β maintenance tool β update component |
8.2 Software UI (Fig 10)
Fig 10 β STEVE UI: β 2D view (left) β‘ 3D view (top right) β’ Panel viewer (bottom center, RI Γ RI-gradient 2D space) β£ Stain picker (bottom right) β€ Button area + sliders + timeline + stain properties.
8.3 Complete acquisition workflow
- Open STEVE β power on 3D Cell Explorer β wait for auto-detection (USB 3.0 must be correctly connected, otherwise Connection failure err.3.3)
- Bright-field selection β use microscope XY+Z knobs to find cell (bright-field is for search only, not for data acquisition)
- Select immersion medium β STEVE prompts for RI (water / DMEM / custom) β wrong = blurry reconstruction
- Single acquisition β click "3D" button or Microscope β Single Shot β select RI β auto-starts (0.5 fps = 2 s result)
- Digital staining β draw ROI in 2D view β pick color β stain appears in panel viewer
- Save β Save button β default .vol format (only STEVE can read)
8.4 Time-lapse
Click "4D" button or Microscope β Time Lapse β 2 dialogs:
- Immersion medium selection
- Acquisition speed + duration (up to 96 h)
Can set auto-stop or manual-stop.
8.5 Digital staining β STEVE's soul
Why important: raw data is a 3D RI volume (looks grayscale); digital staining lets you "color-code different RI values" = label-free multiplexed immunofluorescence.
10 steps (P51-P52):
- Select meaningful slice (drag in 2D view / slider)
- "+" new stain
- Name (optional)
- Pick color
- Draw ROI in 2D view (pixels light up; release mouse β stain range calculated)
- Stain shows in panel viewer (RI Γ RI-gradient space)
- Adjust opacity / edge softness / shape
- Add more stains β repeat
- One acquisition = up to 7 RI digital stains + 3 fluorescence = 10 simultaneous markers
- Stain can be hidden / deleted / recolored
π‘ Digital staining precision depends on sample preparation RI uniformity β see Best Staining Practices (P56 download link).
8.6 Save and export
Fig 16 β Save button β default .vol.
Fig 17 β File β Export β range (single/all/every N frames), data type (RI/fluorescence), format (raw/tiff/obj/png-jpeg).
Notes:
- .vol format = only STEVE can read; for downstream analysis, export tiff/raw/obj
- obj = standard 3D mesh; import to Imaris/Amira/Arivis for advanced rendering
- 4D data export can clip to A-B marker range
9. Troubleshooting π§
P57-P58 error code list:
| Code | Module | Cause + Fix |
|---|---|---|
| err.1.1 | Calibration failed | Too much liquid above sample / rotating arm path blocked / dish wall blocks beam β move to dish center |
| err.1.4 | Calibration failed | Sample absorbs too much (dense dye/thick tissue) β move or contact supplier |
| err.2.1 | Acquisition failed | Position bad β move region / clean MO / check sample prep |
| err.3.1 | Connection failed | USB 2.0 loose / scope not powered β restart PC + replug |
| err.3.2 | Connection failed | USB 3.0 not tightened / not in PC USB 3.0 port |
| err.3.3 | Connection failed | USB 3.0 line detected as USB 2.0 (driver issue) β try other port / install USB 3.0 driver |
| err.3.4 | Connection failed | Scope init failed (rotating arm blocked) β power off β check arm free rotation |
| err.3.6 | Connection failed | Scope needs service β contact supplier |
| err.3.7 | Connection failed (fluo) | Fluorescence module connection failed β check CoolLED power + USB |
| err.3.8 | Connection failed | Firmware update error β contact support |
| err.5.1 | Processing failed | GPU doesn't meet requirement (at least NVIDIA 2 GB discrete VRAM) β update NVIDIA driver |
10. Safety/regulatory essentials β οΈ
P1-P5 β must read, customers often ignore.
- Class I laser (Ξ»=520 nm): harmless to eyes, but no unauthorized opening of cover to access laser path
- CoolLED source (fluo only): may emit UV (depends on wavelength config) β don't look directly + don't cover light guide (fire risk)
- EN/IEC 61010-1:2010 + EN62471:2008 dual certification (light source risk group 3)
- No flammable/toxic solvents for cleaning; ethanol/isopropanol OK (away from ignition, ventilated)
- Not for humid environments
- User responsible for bio-contamination prevention (clean after each use)
- Disconnect power before any maintenance
11. Key resources π
- Nanolive official: http://nanolive.ch
- Nanolive applications: http://nanolive.ch/applications
- Filter config details: https://www.semrock.com/SetDetails.aspx?id=3297
- CoolLED source data: http://www.coolled.com/product-detail/led-wavelengths/
- Chinese SOP (Sample Preparation Protocol): contact supplier
- Chinese digital staining training: contact supplier